A&D MX-50 Thermostat User Manual


 
MS-70 / MX-50 / MF-50 / ML-50 19
1. Turn on the analyzer.
The gram unit (of weighing mode) is displayed.
Enter the Standard Mode
Enter the Standard ModeEnter the Standard Mode
Enter the Standard Mode
2. Press the SELECT key to display an analysis
mode and press the or key to select 5td .
Select ACCURACY
Select ACCURACYSelect ACCURACY
Select ACCURACY
3. Press the SELECT key to select ACCURACY.
When ACCURACY is selected, HI, MID. or LO blinks.
4. Press the or key to select LO of ACCURACY.
Set Drying Temperature at the Sample Pan
Set Drying Temperature at the Sample PanSet Drying Temperature at the Sample Pan
Set Drying Temperature at the Sample Pan
5. Press the SELECT key to select the drying
temperature.
6. Press the or key to set 130 °C.
Store the Parameters and Finish the Operation
Store the Parameters and Finish the OperationStore the Parameters and Finish the Operation
Store the Parameters and Finish the Operation
7. Press the ENTER key to store the parameters.
The weighing mode is automatically displayed.
Put a Sample on the Pan
Put a Sample on the PanPut a Sample on the Pan
Put a Sample on the Pan
8. Put the breeze break ring, pan support, pan handle
and sample pan in order. (With no sample.)
9. Close the heater cover.
10. When displaying a stable value, Press the RESET
key. Avoid mechanical vibration, breeze and
environmental noise during measurement. If it
deviates from zero display, press the RESET key.
Close
Preparation