MS-70 / MX-50 / MF-50 / ML-50 20
11. Open the heater cover. Put in a sample using the
level indicator.
Caution
The sample needs to be more than 0.1 g.
Spread the sample evenly.
12. If the ↑ key is pressed during gram display, analysis
mode, measurement unit and minimum scale value
are displayed. If the ENTER key is pressed, the
sample mass is output.
Start the Measurement
Start the MeasurementStart the Measurement
Start the Measurement
13. Close the heater cover. Press the
START key
after a stable value is displayed.
14. If the SELECT key is pressed during measurement,
other units can be displayed temporarily and its
measurement values can be output.
Caution
Do not press the SELECT key while
sampling data.
Finish the Measurement
Finish the MeasurementFinish the Measurement
Finish the Measurement
15. When the change of moisture content per one
minute (drying rate) reaches the preset termination
value, the measurement is completed and the
buzzer beeps.
16. Open the heater cover and remove the sample
using the pan handle.
17. Press the following key to return to gram display.
ENTER key..... Outputs (Prints) result.
SELECT key ... Returns to the weighing mode.
RESET key ..... Returns to the weighing mode and
displays zero.
18. If the same condition is used, proceed to step 8.
If changing the condition, proceed to step 2.
The sample pans can be washed and reused.
There is the Reference card on the bottom of the analyzer.
Sample needs more than 0.1 g.
Significant sample is indicated
segment.
Range of proper sample
No good Good
Close
Remove sample
Drying temperature
Analysis time
Drying rate
Result