A&D MX-50 Thermostat User Manual


 
MS-70 / MX-50 / MF-50 / ML-50 18
7.
7.7.
7. Measurement Procedure
Measurement ProcedureMeasurement Procedure
Measurement Procedure
7.1.
7.1.7.1.
7.1. Standard Mode Operation
Standard Mode OperationStandard Mode Operation
Standard Mode Operation
The standard mode can obtain the moisture content with settings of ACCURACY
and drying program (heating pattern, drying temperature).
7.1.1.
7.1.1.7.1.1.
7.1.1. ACCURACY
ACCURACYACCURACY
ACCURACY
ACCURACY of measurement can be set either HI, MID. or LO.
The sample quantity is automatically selected by ACCURACY.
The termination value of the analyzing mode is automatically selected by
ACCURACY and minimum scale value of % display.
The analysis mode is the program to finish the measurement when a change of
moisture content per one minute becomes smaller than a preset termination value.
The settings are as follows: Specify an ACCURACY.
ACCURACY
Model Minimum scale HI MID. LO
0.001 % 0.01 %/min 0.02 %/min 0.05 %/min
0.01 % 0.02 %/min 0.05 %/min 0.10 %/min
MS-70
0.1 % 0.10 %/min 0.20 %/min 0.50 %/min
0.01 % 0.02 %/min 0.05 %/min 0.10 %/min
MX-50
0.1 % 0.10 %/min 0.20 %/min
0.05 % 0.05 %/min 0.10 %/min
0.1 %
MF-50
1 %
0.1 %
Preset
Termination
value
ML-50
1 %
0.10 %/min 0.20 %/min
0.50 %/min
Sample quantity 10 g 5 g 1 g
Use Precise result
Quick measurement
7.1.2.
7.1.2.7.1.2.
7.1.2. Operation
Operation Operation
Operation
This operation explanation uses the following example of the MX-50:
Refer to "8.Selection of Measurement Method" for detail.
Input Parameters
Analysis mode ...................................Standard mode
Drying temperature............................ 130 °C
ACCURACY....................................... LO
Sample quantity.................................Approximately 1 g (Automatic selection)
Analysis mode ................................... 0.10 %/min (Automatic selection)
Stored Parameters (Factory Settings for the MX-50)
Heating pattern ..................................Standard drying ( )
Measurement unit ...............................Moisture content is based
on a wet sample
Minimum scale value of % display........0.01 %
Minimum scale value of gram display ... 0.001 g
Data memory function........................ Not used
% MOIST
D/W/D