Teledyne 633 Carbon Monoxide Alarm User Manual


 
Model 633 Aethalometer
®
Black Carbon Monitor User Manual Teledyne Advanced Pollution Instrumentation
68 07645A DCN6170
5.8Calibratingthetapesensor
Tapesensorcalibrationisperformedatthefactoryandshouldonlybeconductedwhen
removingtheMeasurementchamberfrom/totheAethalometer.Thisprocedurerequires
specialcalibrationcylinders:thesmallcylinderhasdiameterof39mm;thelargecylinderhas
adiameterof102mm.
(NOTE:Itisessentialtosecurethecalibrationcylinderswiththetaperetentiondisksduring
calibration).
Tapesensorcalibrationprocedure:
1.IntheOperation/General
menupressStoptostop
measurements,thencarefully
removethesamplingtape.
(RefertoSection3.4 Filter tape
installation).
At this point, remove the tape as instructed in the Change Tape Procedure, then continue as follows:
2.IntheOperation/Advanced
menupress“TapeSenAdj”