3.4Manuallylifting/loweringchamberopticalhead
Incasetheinstrumentisturnedoff,thechambercanbeliftedmanually,asfollows:
(Stopmeasurementsfirst:
Operation>General>Stop).
Notetheretentionpinat
thetopofthechamber
opticalhead.
Grasptheopticalhead
securelyinbothhands
withfingersatbottomand
thumbsontop,pressing
retentionpindownand
raisingopticalheadup.
Figure 11. Manually Lifting the Chamber Optical Head
ATTENTION!Ifthechamberisliftedwhileinstrumentisoperating(STOPbuttonnot
pressed),itisnotpossibletorestartthemeasurementwithoutcyclingtheinstrument
power.(FlipbothpowerswitchestoOFF;thenflipeachtoON).