4 USER INTERFACE, SETTINGS and OPERATION
4.1Userinterfaceandsettings
Therearefour1
st
levelmenutabsinthetouch‐screenuserinterface:HOME,OPERATION,
DATA,andABOUT(Figure 12),describedasfollows:
HOMEwiththefollowingparameters:
o BC‐themeasuredvaluesforBlackCarbon(measuredat880nm)
o UVPM‐UVparticulatematter(calculatedat370nm)
o REPORTEDFLOW‐measuredflow(inLPM)
o TIMEBASE‐timebasesetting(inseconds)
o TAPEADV.LEFT‐theamountofremainingtape(in)
o STATUS–instrumentstatus:green(allOK),yellow(checkstatus),andred
(stopped)withaStatusCondition(see,4.2 Instrument Status),
o BottomofHOMEscreenshowsdateandtime.
Note:TheBCandUVPMvalueswilltypicallybesimilar,butnotexactlythesame.
Ifaromaticcompoundsarepresentinthesampledair(suchaswhensampling
freshwoodsmoke,forexample),theUVPMconcentrationwillexceedtheBCmass
concentrationvaluesignificantly,dependingupontheamountandtypeoforganic
materialpresent.
OPERATIONwithfoursub‐tabs:GENERAL,ADVANCED,LOGandMANUAL.
GENERALwhereonecanchangethesettings:
o Flow(LPM)(seeNotebelowonflowreportingstandard)
o TimeBase(seconds)
o selectoneofthreeradiobuttons:
TAATNmax‐maximumattenuationatwhichtheAethalometeradvances
tape
TAINT‐thetimeintervalatwhichtheAethalometeradvancestape
(hours)
TATime‐thetimeatwhichtheAethalometeradvancestape
o Start,Stop,Stability,Cleanair,andChangeTapebuttons
ADVANCEDwithallparameterswhichcanbesetintheAethalometer.
LOGwiththelastoperationalreportsofstatus,parameterchanges,datadownload.
MANUALwith basiccommands, tooperate hardware(solenoids,pump, chamber,
TA)
Note:flowreportingstandard:Thesearethepressureandtemperaturewhichthe
instrumentusestoreportflowandsamplevolume.Bychoosingtoreportmass
flow,(101325Pa,21.11C)areusedtoconvertthemeasuredmassflowtoreported
volumetricflowattheseconditions(thesearevaluesusedbytheflowmeterinthe
Model633).Thesevaluescanbesubstitutedforanyother(pressure,Temp)values,