Teledyne Advanced Pollution Instrumentation Model 633 Aethalometer
®
Black Carbon Monitor User Manual
07645A DCN6170 59
Manualflowcalibrationprocedure:
10. Inthedisplay’sblank
FLOWfieldentertheflow
valuefromyourflow‐meter
measuredinorconverted
tomilliliters(thiswillbe
approximately3000;do
NOTchangethepressure
ortemperaturesettings),
andselectOK.
11. Waitforthepumptorun
forseveralsecondsto
achieveastableflow.
12. Inthedisplay’sblank
FLOWfieldentertheflow
valuefromyourflow‐meter
measuredinorconverted
tomilliliters(thiswillbe
approximately5000;do
NOTchangethepressure
ortemperaturesettings),
andselectOK.