Infinity M30 Burner User Manual


 
ERROR CODES
EERRRROORR
CCOODDEE
DDEESSCCRRIIPPTTIIOONN
PPRROOBBAABBLLEE CCAAUUSSEE
SHORTED THER-
MOCOUPLE OR
DEFECTIVE HEATER
PLATES
Occurs if the thermocouple wires are shorted or the heater plates are
defective causing Infinity to achieve a HEAT RATE of less than 6% of the
maximum attainable HEAT RATE with full power applied to the heater
plates for 5 minutes. This error will also occur if a program is running
and the door is kept open for more than 5 minutes or the triac or the
Main PC board developed a problem during a burnout process.
Er 6
THERMAL
RUNAWAY
Er 7
INVALID ENTRY
ERROR WHILE A
PROGRAM IS
RUNNING
Flashing
99:99
when
NIGHT
TIME
(DELAY
START) is
pressed
during a
program
Occurs when the HEAT RATE in STAGE 2, 3 or 4 is set to “COOL” but
the corresponding TEMP is entered to heat to a higher temperature
than the previous stage.
Occurs when the temperature has exceeded 2015°F (1102°C) for 1
minute if the highest temperature programmed was less than 2000°F
(1093°C). This also occurs when the temperature has exceeded
2050°F (1121°C) for 1 minute if the highest temperature programmed
was 2000°F (1093°C) or higher.
18