Programming the Status System 15Chapter 2
Status System Programming Examples
This section contains two example programs that use the status system and
common commands to monitor when data is available from an instrument
and when an error has occurred. Both programming examples are written in
C and use the Standard Instrument Control Library (SICL) for I/O
operations. The example programs use SCPI (Standard Commands for
Programmable Instruments) commands to communicate with the status
system. Thus, the instruments must either be message-based or have a SCPI
interpreter, such as an HP E1405/06 Command Module or the SICL iscpi
interface.
Handling SRQs The following is a general procedure for handling SRQs:
Define the SRQ handler to do the following:
-- Read the Status Byte using ireadstb. ireadstb returns the RQS
(request for service) bit in bit 6 of the status byte. After issuing a
ireadstb, RQS is cleared indicating that the Service Request is being
acknowledged. A new SRQ will not be issued unless RQS is
cleared. Using *STB? will return the Master State Summary in bit 6
and does not affect RQS, therefore this should not be used in a SRQ
handler.
-- Check the status byte to determine which status group(s) requires
service.
-- For each status group that requires service, read the event register of
that status group to determine what caused the SRQ to be generated.
It is necessary to clear the event register so that if a new event
occurs a new SRQ will be generated.
-- Take some action after determining which event caused the SRQ.
The action taken is determined by evaluating the contents of the
event register.
Enable SRQ Handler in SICL with ionsrq.
Make sure that all the Enable Masks in all the status enable registers
are set to the proper values to propagate the summary bit(s) to the
status byte. An SRQ is only generated if the MSS (Master State
Summary) bit in the status byte is set.
Using MAV to
Determine When
Data is Available
The following example program sets up an SRQ handler to be called when
there is data in the output queue. The program then prompts for SCPI
commands. If the SCPI command results in data in the output queue (such
as a query command), then the SRQ handler is called and the data is printed.
The following summarizes the procedure used: