GE Z-6000 Water System User Manual


 
Z Series RO System by GE Osmonics
4.4.1 Formaldehyde Disinfection Preparation
WARNING:
Always wear rubber gloves, apron and facemask. Use proper ventilation during
disinfection and rinse-out. A face respirator with an organic cartridge to remove
formaldehyde fumes may be worn. DO NOT BREATHE FUMES OR ALLOW
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES. Follow all
Material Safety Data Sheet (MSDS) precautions and OSHA Standards.
WARNING:
Label machine with appropriate warning signs such as "Do Not Use / Contains
Disinfectant" to prevent use of the RO until it is properly rinsed.
CAUTION:
New or replacement thin film membranes must be rinsed to drain with normal pressures
for a minimum of six (6) hours before first exposure to formaldehyde, or irreversible
damage may occur.
Disinfecting Materials
Description GE Osmonics Part Number
Disinfection/Cleaning Kit 12214
Formaldehyde, 37% Solution Not Provided by GE Osmonics
Formaldehyde Test Kit Not Provided by GE Osmonics
“DO NOT USE” Warning Labels Not Provided by GE Osmonics
Cleaning/Disinfect Log Sheets Reference Tech Notes 143, 144, 145
Operation Log Sheet Reference Tech Note 142
Body Protective Wear Not Provided by GE Osmonics
1238339b – 16Oct03 4 - 16 Routine Maintenance