3 Specifications and Environmental
Conditions
3.1 Specifications
Operating Range 300°C to 1100°C
Stability 300°C ±0.05°C
500°C ±0.05°C
700°C ±0.1°C
1000°C ±0.1°C
1100°C ±0.1°C
Uniformity 300°C ±0.05°C
500°C ±0.08°C
700°C ±0.2°C
1000°C ±0.25°C
1100°C ±0.3°C
Stabilization Time Typically 2 hours midrange, slower at the low temperature end (4
hours), faster at the high temperature end
Heater Power 3700 Watts High
Power Requirements 230 VAC (±10%), 50/60 Hz, 20 A
Outside Dimensions 18” H x 14.25”W x 26”D (457mm x 362mm x 660mm)
Weight 72.5 lbs
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
•
temperature range: 5 - 40°C (41 - 104°F)
•
ambient relative humidity: 15 - 50%
•
pressure: 75kPa - 106kPa
•
mains voltage within ± 10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitude less than 2000 meters
•
indoor use only
3.3 Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be
free from defects in material and workmanship under normal use and service
11
3 Specifications and Environmental Conditions
Specifications