MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)
11
Non-heater wire operation:
Invasive Mode: Airway temperature < 29.5 °C causes an audible
and visible alarm.
Non-invasive Mode: Airway temperature < 26.0 °C causes an
audible and visible alarm.
Sound Pressure Level: Alarms exceed 50 dBA @ 1 m.
3.4 Performance
Invasive Mode: Flow up to 60 LPM, humidity output >33 mg/L
Non-Invasive Mode: Flow up to 120 LPM, humidity output >10 mg/L
NOTE: Performance results with RT100 breathing circuit
Maximum System
Operating Pressure:
20 kPa, gas leakage at max. pressure <100 mL/minute.
Warm-up time: Less than 30 minutes.
Recommended ambient
temperature range:
18 to 26 °C
CAUTION:
If operating outside the recommended ambient temperature range, consult your
local Fisher & Paykel Healthcare representative or refer section 4.1.5.
3.5 Transport and Storage Environmental Conditions
Transport Conditions: -10 to +50 °C
Storage Conditions: -10 to +50 °C
3.6 Standards and Approvals
AS/NZS 3200.1.0, CAN/CSA 22.2 No.601.1, UL 60601-1, IEC 60601-1, EN 60601-1