164 Set calibration parameters
Section 5: Startup and meter configuration Reference, Installation, and Operations Manual
June 2013 3-9000-743 Rev S
2. Choose a Flow Calibration Method:
• Polynomial
• Piece-wise linear
• None
3. Enter the appropriate forward and reverse meter factors and click the
Write button to
send the factors to the meter.
4. Click the
Cancel button to close the dialog box. If you made changes, click the Write
button to apply the changes.
After the meter is configured, set S-2 Position 4 (on the CPU Board) back to the open
position to write-protect the configuration.