Compur Gas Detector Carbon Monoxide Alarm User Manual


 
-
34
-
??The label on the sensor shows the span gas to be used (e.g.: Butane) and a response
factor. Multiply the concentration of your span gas [% UEG] with the resonse factor
shwon on the sensor label. Perform a gas calibration as mentioned in 4.2.1. Enter the
result of the multiplication as your span gas concentration.
4.3 Changing the Alarm Levels
Compur Ex plus features two adjustable alarm levels. Standard factory setting is 25 % and
50 % LEL.
Alarm threshold A1 may not be higher than alarm level A2 and alarm threshold A2 may not
be lower than A1. If you want to adjust A1 to more than actual A2, you first need to increase
A2.
Adjusting alarm level A1
??Turn the function switch to “TEST”- position. Instrument will perform an automatic
selftest.
??Press “Cal” using the keyboard pin. The display will show “000” indicating the automatic
zero adjust.
??When the zero adjust is finished, display will show “GCAL”.
??Press “Fct”. Display indicates “A1”.
??Press “Cal” once. Display will show current A1 adjustment.
??Press and hold “Cal”. Display will increase until it has reached A2 and than start again at
zero.
??If display has reached the desired value for A1, stop pressing “Cal” and press “Fct”.
Display shows “Stor” indicating that the new value has been stored in memory.
??After storage the instrument will show the actual measured value.
Adjusting alarm level A2
??Follow the procedure as described in A1 adjustment until the display shows “GCAL”.
??Press “Fct” twice. Display will show “A2”.
??Press “Cal” once. Display will show current A2 adjustment.
??Press and hold “Cal”. Display will increase until it has reached 100 % LEL and than start
again at A1.
??If display has reached the desired value for A2, stop pressing “Cal” and press Fct”.
Display shows “Stor” indicating that the new value has been stored in memory.
??After storage the instrument will show the actual measured value.