Signal Output
Zeroing signal output
69
The function of
ZERO
is to subtract a constant background signal from
the detector signal. Background signal sources include the detector itself
(background level depending upon detector type), column bleed, or
contaminants in supply gas(es).
Displaying current
ZERO
setpoint
Current
ZERO
setpoint value is displayed by pressing the appropriate
signal channel key (
SIG 1
or
SIG 2
), followed by
ZERO
(or simply press
ZERO
alone, if the desired signal channel is already displayed). Typical
displays are shown in Figure 5•1.
ACTUAL SETPOINT
SIG 1 ZERO 104.5
ACTUAL SETPOINT
SIG 2 NOT INSTALLED
ACTUAL SETPOINT
ZERO 1 104.5 OFF
ACTUAL SETPOINT
ZERO LIMIT = 83000
Typical Displays
ZERO
Figure 5-1
Table 5•1lists conversions between detector units and the scale used for
ZERO
setpoint values.
FID & NPD 1.0 pA (1.0 10
-12
A)
TCD 25
V (2.5 10
-5
V)
ECD 10 Hz
Detector
Conversion
(1.0 unit = )
ZERO
Table 5-1. Conversions from Detector Units to Setpoint Val-
ues
ZERO