Infinity L30 Burner User Manual


 
20
1. If a power failure occurs, the Infinity memorizes the conditions prior to the loss of power. When the power
returns, the Infinity returns to the proper point in the program.
2. When power is returned, the HR:MIN light flashes indicating that a power failure has occurred. It
continues to flash until START / STOP is pressed.
NOTE: The HR:MIN light flashes if the power switch is turned off and on while a
program is running and START / STOP was not pressed. It will not flash if the
power switch was turned off or a power failure occurred when the PROGRAM
READY light was on.
REPLACEMENT OF DOOR INSULATION AND SPRINGS
Infinity M PN 15722 Door Insulation
Infinity L PN 15712 Door Insulation
Infinity M and L PN 33997 Door Spring and Hook Assembly Kit
1. Open, locate and remove the two screws on the door closest to the door hinges which hold the retainer
strip in place. Remove the retainer strip.
2. Remove the one piece door insulation by sliding it toward the rear of the furnace and slightly lifting.
3. To replace the springs, remove each spring from the hook which holds it in place. Remove both the
hook and spring.
4. To reinstall the new door insulation or springs and hooks, reverse the above procedure. If installing
springs and hooks, add grease to junction of spring and hook and spring and hinge junction.
POWER FAILURE
SERVICE
6. When the tablet begins to melt or liquify at the edges, immediately press and hold.
Then press ENTER/REVIEW. Your Infinity is now calibrated. Three “beeps” sound and
“CAL” appears on the main display.
°
Heating
Chamber
Calibration
Tablet
Ceramic
Tray
Casting
Ring