Honeywell MicroDock II Carbon Monoxide Alarm User Manual


 
MicroDock II
Quick Reference Guide
12
GasAlertMicroClip
GasAlertMax XT
The detector can be configured prior to performing a bump test
or calibration to
change user options,
sensor options, and
program options.
The detector can also be configured to disallow bump tests or
calibrations unless performed with the MicroDock II. For more
information, refer to Configuring Detectors in the Fleet Manager
II Technical Reference Guide.
Charging the Battery Pack (Optional)
a Warning
There is a maximum of six charging docking modules can
be installed on the MicroDock II base station (six charging
modules plus four non-charging modules for a maximum
of ten docking modules per station).
To charge successfully, the temperature must be between
50°F to 95°F (10°C to 35°C). Charge the battery
immediately when the detector emits a low battery alarm.
a
Caution
The charger is for indoor use only.
Note
The detector will not charge while bump tests or
calibrations are being performed. If the battery is low,
charge the detector for 30 minutes, then begin a test.
When the test is complete resume charging the
battery.
When charging a new battery for the first time, ensure the bat-
tery obtains a full charge. For more information, refer to the
detector’s user manual or operator’s manual.
Note
Up to six GasAlertMicro and GasAlertMicro 5/PID/IR
detectors can charge simultaneously.
Up to four GasAlertMicroClip and GasAlertMax XT
can charge simultaneously.
For more information, refer to the MicroDock II Base
Station User Manual.
To charge the battery pack, complete the following:
1. Connect the charger adaptor cord into the CHARGE
port on the station, and then plug into an AC outlet.
The CHARGE LED briefly lights red then green dur-
ing the self-test. The LED then powers off (not appli-
cable to the GasAlertMicroClip and GasAlertMax XT
modules).
2. Deactivate the detector. Insert the detector into the
charging cradle.
3. The CHARGE LED lights red (all charging modules).