Emerson 3200 Heat Pump User Manual


 
60 cm H
2
O (44 mm Hg)
60 cm H
2
O (44 mm Hg)
Conforms to UL STD 2601-1, certified to CAN/CSA - STD C22.2 No. 601.1-M90, 0413
3.3 liters/second with inhale flow set to minimum; if set to maximum inhalation, the flow is
the same as the exhalation flow
10 liters/second; actual flow depends on maximum pressure and on airway resistance
-70 to 0 to +70 cm H
2
O; accuracy +/-2% of full scale
Models CA-3000, CA-3200: Automatic and Manual Timing
Models CM-3000, CM-3200: Manual Timing
CA-3000, CA-3200: Automatic mode, 0 to 5 seconds
CM-3000, CM-3200: User variable
Yes - connects to ambient
Two-stage centrifugal blower with AC/DC universal motor
CA-3000, CM-3000: 100-120 VAC, 60Hz
CA-3200, CM-3200: 220-240 VAC, 50Hz
CA-3000, CM-3000: 300 VA
CA-3200, CM-3200: 600 VA
292mm x 279mm x 419mm (HxWxD)
(11.5” x 11” x 16.5”)
CA-3000, CA-3200: 11 kg (24 lbs)
CM-3000, CM-3200: 9.3kg (20.6 lbs)
Operating temperature range: +10 to +40°C (+50 to +104°F)
Storage temperature range: -20 to +50°C (-4 to +122°F)
Operating humidity range: 30% to 75%
Storage humidity range: 10% to 90%
SECTION 3
S PECIFICATIONS
Maximum Positive Pressure:
Maximum Negative Pressure:
Standards:
Maximum Inhalation Flow:
Maximum Exhalation Flow Capacity:
Pressure Gauge:
Mode of Operation:
Inhalation, Exhalation, Pause Times:
“Off” Position:
Blower Type:
Input Voltage:
Input Power:
Dimensions:
Weight:
Temperature Specifications:
Humidity Specifications:
6
92804 3/28/01 8:35 AM Page 8