Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-13
Philips Medical Systems
4
Voice prompts:
ANALYZING INTERRUPTED
or
CANNOT ANALYZE
or
STOP ALL MOTION
• The patient is being moved or jostled.
• Radio or electrical sources are
interfering with ECG analysis.
• The environment is dry and movement
around the patient is causing static
electricity to interfere with ECG
analysis.
• Stop CPR; do not touch the
patient. Minimize patient motion. If
the patient is being transported,
stop the vehicle if needed.
• Check for possible causes of
radio and electrical interference
and remove them from the area.
• Responders and bystanders
should minimize motion,
particularly in dry environments
that can generate static electricity.
Voice and screen prompts:
NO SHOCK DELIVERED
The patient impedance is not appropriate
for the FR2+ to deliver a shock.
• Make sure the defibrillator pads
are correctly positioned on the
patient according to the diagram
on the back of the pads.
• Make sure the defibrillator pads
connector is completely inserted
in the connector socket.
• Press the defibrillator pads firmly
to the patient's chest.
• Replace the defibrillator pads if
necessary.
Voice prompt:
SHOCK BUTTON NOT
PRESSED
Shock has been advised but not delivered
within 30 seconds. (FR2+ has been
disarmed.)
• When next prompted, press the
Shock button to deliver shock.
symptom possible cause recommended action