Omron Healthcare U22 Air Cleaner User Manual


 
1918
Troubleshooting
The nebulisation rate
is extremely low
Trouble Possible cause Corrective action
Batteries are low (the orange
battery low indicator flashes).
Replace the batteries with new
ones. Refer to page 8.
The mesh is stained or clogged. 1. Clean the mesh by boiling. Refer
to page 15.
2. Replace the mesh cap with a new
one. Refer to page 10.
The green power indi-
cator does not turn on
and the unit does not
nebulise.
The batteries are low. Replace the batteries with new
ones. Refer to page 8.
The capacity of rechargeable
batteries is low.
Charge the batteries with a com-
mercially available recharger.
The batteries are inserted in the
wrong direction.
Re-insert the batteries in the correct
direction. Refer to page 8.
The AC adapter is not correctly
connected to the main unit.
Remove the AC adapter from the
main unit and connect it correctly.
Remove the medication container
from the main unit and install cor-
rectly. Refer to page 11.
The medication container is not
installed correctly to the main unit.
Tilt the main unit in order to
immerse the vibrator in the medica-
tion.
The vibrator is not immursed in the
medication.The green power
indicator turns on but
the unit does not
nebulise.
The medication container is not
filled with medication.
Fill with medication. Refer to page 11
The mesh is broken.
Replace with a new one. Refer to
page 10.
The mesh cap is a consumeable part.
Omron recommends to exchange the mesh cap after approximately 1 year.
In case the unit nebulises poorly even after boiling in water (refer to Disinfection by boiling), replace
the mesh cap.
Remove the mesh cap and install a new one as described on page 10 (section How to assemble the
unit and fill the medication container).
How to replace the mesh cap