Omega Vehicle Security FPU5-MT-220 Heat Pump User Manual


 
6.3 The TIME Mode
Figure 6-5 shows the overall block diagram of the TIME Mode in the
RUNNING Display Mode.
Figure 6-5. Block Diagram: Pumping in the TIME Mode
In the TIME Mode, you pump fluid at a certain motor speed or flow rate, for a
particular amount of time. The pump motor uses the RPM or FLOW RATE
setting. It measures the volume of fluid dispensed, but ignores the TOTAL
VOLUME setting.
6.3.1 Pumping in the TIME Mode
To pump fluid in the TIME Mode, do the following:
1. Set up the pump motor for pumping according to Chapter 3.
2. Toggle the ON/OFF switch to turn on the unit’s power. The display is in the
STOPPED Display Mode. Refer to Figure 6-5.
3. Using the procedure outlined in Sections 6.1.1. and 6.2.1, adjust either the
RPM or FLOW RATE setting.
STOPPED
DISPLAY
MODE
OR
TUBE ID SETTING
TO BE USED IN
PUMPING
OR
CURRENT
TUBE ID
SETTING
CURRENT
RPM
SETTING
OR
OR
FLOW
RATE
CURRENT
FLOW RATE
SETTING
FLOW RATE
SETTING
TO BE USED IN
PUMPING
(Optional)
CALIBRATE
TUBE ID
RPM SETTING TO BE
USED IN PUMPING
(Optional)
CALIBRATE
RPM
TIME
START
STOP
RUNNING
DISPLAY
MODE
and
TIME ELAPSED
DURING PUMPING
CURRENT
TIME SETTING
TIME SETTING
TO BE USED IN
PUMPING
TIME
START
STOP
(Forced Stop)
(Time Runs Out)
(Refer to Section 5.5)
(Refer to Section 5.5)
Operating the Pump Motor in the RUNNING Display Mode
6
6-11