Figure 6-4. FLOW RATE Ranges (cont’d)
6. Once the display shows the FLOW RATE setting you desire, press the
START/STOP key.
As the fluid begins to dispense, the non-flashing display shows the calculated
fluid flow rate.
Refer to Section 6.5 to learn about how to prevent tubing from moving
through the pump.
7. To stop pumping in the FLOW RATE Mode, simply press the START/STOP
key again. The pump stops running, and the display returns to the STOPPED
Display Mode.
Table 6-2 shows how other keys and buttons can be used while pumping in the
FLOW RATE mode.
xxx mL/min
START
STOP
ACTION
Pump(s) starts;
press again pump(s)
stops
Display
Range (7.0 mm tube ID)
37 mL/min
38 mL/min
.
1999 mL/min
2000 mL/min
.
2186 mL/min
2187 mL/min
Range (7.0 mm tube ID)
2187 mL/min
2186 mL/min
.
2000 mL/min
1999 mL/min
.
38 mL/min
37 mL/min
Range (8.0 mm tube ID)
39 mL/min
40 mL/min
.
1999 mL/min
2000 mL/min
.
2315 mL/min
2316 mL/min
Range (8.0 mm tube ID)
2316 mL/min
2315 mL/min
.
2000 mL/min
1999 mL/min
.
40 mL/min
39 mL/min
Operating the Pump Motor in the RUNNING Display Mode
6
6-8
NOTE