8544install.fm Installation
Section 3 Installation
DANGER
Only qualified personnel should conduct the installation tasks described in this
section of the manual
3.1 Unpacking the Instrument
Remove the sampler from the shipping carton and inspect it for any damage.
Contact Hach Customer Service at 1-800-227-4224 if any items are missing
or damaged.
3.2 Selecting the Installation Site
The sampler is completely self-contained and may be located indoors or
outdoors without protection, between 0 to 50 °C (32 to 120 °F). Follow the
guidelines below and
Figure 3 to allow complete drainage of the intake line
and prevent cross-contamination between samples.
• Install the sampler as close to the source as site conditions permit. This
will increase pump tube life and optimize overall sampler performance.
• Mount the sampler to a stable flat surface using the wall mounting
bracket, Cat. No. 3596 (
Figure 4).
• Install the sampler above the sample source, with the intake tubing
sloping downward to the sample.
• Make sure the tubing is free of kinks or loops.
Figure 3 Placement of the Sampler
1. Slope tubing down to source (no loops, kinks, or excess tubing).
2. Locate strainer in an area of turbulent and well-mixed flow.
1
2
Flow