Specifications
Dual Input Meter
4
61
Measurement Current..........................0.5 mA to 50 nA
decreases with increasing ranges
Open Circuit Voltage .............................................. <4V
Continuity (CONT)
Beep .....................................<(30 ±5) in 50 range
Measurement Current........................................ 0.5 mA
Detection of shorts of .......................................... 1 ms
Diode
Measurement Voltage:
@0.5 mA........................................................... >2.8V
@open circuit....................................................... <4V
Accuracy..............................................±(2% +5 counts)
Measurement Current........................................ 0.5 mA
Polarity......................................+ on input A, - on COM
Capacitance (CAP)
Ranges .................50 nF, 500 nF, 5 µF, 50 µF, 500 µF
Accuracy............................................±(2% +10 counts)
Full Scale Reading .................................... 5000 counts
Measurement Current............................5 µA to 0.5 mA
increases with increasing ranges
Dual slope integrating measurement with parasitic
serial and parallel resistance cancellation.
Advanced Meter Functions
Zero Set
Set actual value to reference
Fast/Normal/Smooth
Meter settling time Fast: 1s @ 1 µs to 10 ms/div.
Meter settling time Normal: 2s @ 1 µs to 10 ms/div.
Meter settling time Smooth: 10s @ 1 µs to 10 ms/div.
Touch Hold (on A)
Captures and freezes a stable measurement result.
Beeps when stable. Touch Hold works on the main
meter reading, with thresholds of 1 Vpp for AC signals
and 100 mV for DC signals.
TrendPlot
Graphs meter readings of the Min and Max values from
15 s/div (120 seconds) to 2 days/div (16 days) with time
and date stamp. Automatic vertical scaling and time
compression.
Displays the actual and Minimum, Maximum, or average
(AVG) reading.
Fixed Decimal Point
Possible by using attenuation keys.