Fluke 80K-15 Air Cleaner User Manual


 
®
PN 690385 May 1998
©1998 Fluke Corporation. All rights reserved. Printed in U.S.A.
Performance Test
Verify the accuracy of the probe by measuring a 15 kV dc
(± 0.25 %) voltage source. When used with a compatible
dc voltmeter, the probe should measure the source with
± 1% accuracy. No calibration adjustments are needed.
Warranty
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material
and workmanship for one year from the date of
purchase. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect,
misuse or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the
warranty period, send your defective tester to the
nearest Fluke Authorized Service Center with a
description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO
OTHER WARRANTIES, SUCH AS FITNESS FOR A
PARTICULAR PURPOSE, ARE EXPRESSED OR
IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL,
INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES OR LOSSES, ARISING FROM ANY CAUSE
OR THEORY. Since some states or countries do not
allow the exclusion or limitation of an implied warranty or
of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA. 5602 B.D. Eindhoven
98206-9090 The Netherlands
10/96
Contacting Fluke
For application or operation assistance, or information
on Fluke products, call:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com.
80K-15
Electronic Air Cleaner Probe
Instruction Sheet
Introduction
The 80K-15 Electronic Air Cleaner Probe (hereafter referred
to as “the probe”) is a low energy probe that extends the
voltage measuring capability of an ac/dc voltmeter or
multimeter.
Properly connected, the probe is rated to 15 kV Overvoltage
Category I. The use of the probe is restricted to low energy
applications, which are energy limited circuits like
electrostatic air cleaners.
Do not use this probe to
measure high voltages on power distribution systems.
The probe is a precision 1000 : 1 voltage divider formed by
two matched resistors. The unusually high input impedance
offered by these resistors minimizes circuit loading and
thereby optimizes measurement accuracy.
W
Read First: Safety Information
WWarning
To avoid possible personal injury or electrical shock:
Use the probe within its voltage ratings and in dry
conditions (no condensation is present).
Be familiar with and exercise all possible high
voltage safety practices.
DO NOT use this probe to measure high voltage on
power distribution systems.
Always hold the probe by its gray handle. Never
allow the probe tip or the yellow part of the probe
to make body contact.
Before making a measurement, make sure that the
side of the probe plug with the GND tab is plugged
into the low (COM) terminal of the meter.
Attach the grounding lead to earth ground.
Use a meter with an impedance of 10 M ( ± 1% ).
Using a meter with an input impedance less than
10 M reduces the accuracy of the reading by
causing it to be lower than the voltage present.
DO NOT use the probe with a Fluke multimeter that
is in the kChek or Automatic Selection mode (i.e.
Fluke 16, Fluke 12B etc.). The input impedance in
these modes is very low.
Make sure the probe tip is firmly attached to the
body of the probe.
DO NOT use the probe around explosive gas,
vapor, or dust.
DO NOT use the probe if it operates abnormally.
Protection may be impaired. When in doubt have
the probe serviced.