Bacharach 50 Carbon Monoxide Alarm User Manual


 
- 2 -
Figure 1. Snifit 50
Specifications
Gas Monitored ............Carbon Monoxide (CO)
Range ..........................0 to 1999 ppm
Resolution ...................1 ppm
Accuracy ......................± (5% of reading + 5 ppm)
Response .....................30 sec. to 90% of final value
Power...........................9V battery
Battery Life ................1500 hours minimum
Temperature Range ....32 to 104 °F (0 to 40 °C)
Case Material .............High impact plastic
Weight .........................0.25 lb (0.11 kg) without battery
Dimensions .................5.5" L x 2.0" W x 1.6" H
(140 x 51 x 41 mm)
Operation
Turning the Snifit
On/Off
If not already done,
install a 9V battery as
described under
Battery Installation.
Turn on the instru-
ment by pressing its
Power On button.
Turn it off by pressing
Power Off (or the
instrument will
automatically shut
itself off after approxi-
mately 35 minutes).
Zeroing the Sensor
Turn on the Snifit and allow the displayed reading to
stabilize before proceeding – approximately 30 sec-
onds. Then with the instrument sampling fresh air
M ONOXIDE ANALYZER
ppm CO
CARBON
S n ifit
m odel 50
Backlight
On
Off
Span Adjust
-P ow er-
ZERO
ADJUST
SPAN
ADJUST
POT.
BACKLIGHT
ON/OFF
POW ER ON
POW ER OFF
DISPLAY
LO BAT
LO W
BATTERY
IN D IC A T IO N