
20
Measurement
3) Ending the Measurement
About Auto-Power-Off function
By connecting the instrument to a personal computer and using
the optional software, the time to be elapsed from power-on
(start of measurement) to power-off (end of measurement) can
be set as the auto-power-off timer.
If an auto-power-off timer is set, the measurement is nished
automatically in the time after power-on.
Regarding the memory function
●
The memory built into the PULSOX-300i will continue to hold
data even if the battery is removed.
●
A maximum of 300 hours of measurement data can be stored; e.g.,
10 hours of measurement data can be stored for 30 persons.
Storage for measurement data is set up in 30-minute blocks,
so even if the duration of measurement data is less than 30
minutes, it will be treated as a 30-minute block. For this reason, if
measurement data with durations of shorter than 30 minutes are
stored, the actual measurement duration will be different from the
stored duration, and the maximum storable time will be shorter
than 300 hours.
●
The maximum storable number of data les is 399 les. After that,
the No. 400 data le will be stored as data le No. 1, overwriting
the previous No. 1 data le.
Press the Power button for a
long time, the measurement is
nished and the Power is off.