Warning
Message
Fault Condition
Possible Causes
on wheel rotation.
Failed Sync/Demod Board
If Status LED’s on the Sync/Demod Board ARE
flashing the cause is most likely an a problem with
the analyzer’s main power:
Intermittent loss of the power supply too short
to cause a SYSTEM RESET warning;
Frequency problem with the AC Mains.
SYSTEM RESET
The computer has
rebooted.
This message occurs at power on.
If it is confirmed that power has not been
interrupted:
Failed +5 VDC power,
Fatal Error caused software to restart
Loose connector/wiring
WHEEL TEMP
WARNING
The filter wheel
temperature is
controlled at 68 ± 5
°
C
Blocked Cooling Vents below GFC Assembly
Analyzer’s Top Cover removed
Wheel Heater
Wheel Temperature Sensor
Relay controlling the Wheel Heater
Entire Relay Board
I
2
C Bus
APPENDIX B – GFC7000EU Spare Parts List
NOTE
Use of replacement parts other than those supplied by API may result in
non-compliance with European standard EN 61010-1.