ResMed s8 Humidifier User Manual


 
39
ENGLISH
SYSTEM SPECIFICATIONS
SYSTEM SPECIFICATIONS
S8 AUTOSET VANTAGE (NO ACCESSORIES)
Performance: Operating pressure range: 4 to 20 cm H
2
O
Dimensions (H x W x D): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power Supply
Input range for S8 AUTOSET VANTAGE with HUMIDAIRE 3i: 100–240 V, 50/60 Hz;
110 V, 400 Hz; 2.5 A <140 VA (110 W) (maximum power consumption).
Instantaneous peak power consumption <340 VA.
Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed
DC-12 converter for DC input.
Actual power consumption will vary, depending on factors such as the use of
accessories, height above sea level, and ambient temperature. Power consumption
values for typical treatment conditions (using a breathing machine set to 0.5 L and
15 breaths/min with 6’ 6” (2 m) air tubing, and an
ULTRA MIRAGE mask) are given in the
following table
Power outlet for HumidAire 3i
Outlet range: 100–240 V, 50/60 Hz; 110 V, 400 Hz; 2.0 A
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6’6” (2 m)
Air Outlet: The 22 mm conical air outlet complies with EN 1281-1
IEC 60601-1 Classifications: Class II (double insulation), Type CF
Noise level: Radiated sound pressure is measured at <30 dBA at 10 cm H
2
O and
1 m (3’3.37”) according to ISO17510
ACCESSORIES
DATA CARD
Dimensions (L x W x D): 2.56” x 0.87” x 0.03” (65.04 mm x 22.15 mm x 0.85 mm)
Device
Pressure
(cm H
2
O)
Power
(VA)
Power
(W)
S8 AutoSet Vantage 5 18 7.6
10 23 11.6
15 30 15.9
20 39 20.7
33887r2.book Page 39 Thursday, June 30, 2005 1:36 PM