Leica MZ16 FA Indoor Furnishings User Manual


 
Supply units
The 100W mercury burners are powered by the ebq 100 supply
unit.
Read the separate instruction manual provided with this
supply unit.
The supply unit for 50W mercury burners
is described in this manual.
For the ebq supply unit: only fuses of the specified type
with the listed rated current may be used. Using other
fuses or bridging the fuse holder is not permitted. Danger
of fire!
Always unplug the power plug of the supply unit from the power
supply:
when installing and disassembling the lamp housing
before opening the lamp housing
when replacing the high-pressure mercury burner and
other parts, such as the heat-absorbing filter or the
collector
during maintenance work on the supply unit.
After switching on the supply unit, new mercury burners
must be adjusted immediately.
There is a danger of overheating for mercury burners that
are not adjusted.
Danger of explosion.
Notice of dangers
Never open the lamp housing when the lamp is switched on.
Eye hazard.
Never open the lamp housing when it is energized.
Disconnect the power supply.
Disconnect the lamp housing from the power supply when
assembling it. Disconnect the power plug and power supply.
Danger of explosion, UV radiation, danger of blinding.
Before opening the lamp housing, allow it to cool off for at
least 15 min.
Danger of explosion, danger of burns.
Never cover the air duct on the lamp housing.
Danger of fire!
5
Leica lamp housing and mercury burner