51 & 52 Series II
Replacement Parts and Accessories
15
Electrical (cont.)
Measurement
Accuracy, T1,
T2, or T1-T2
(Model 52)
J-, K-, T-, and E-type: ±[0.05 % of
reading + 0.3
o
C (0.5
o
F)]
[below −100
o
C (−148
o
F): add 0.15 %
of reading for J-, K-, E-, and N-type;
and 0.45 % of reading for T-type]
Temperature
Coefficient
0.01 % of reading + 0.03
o
C per
o
C
(0.05
o
F per
o
F) outside the
specified +18
o
C to 28
o
C (+64
o
F
to +82
o
F) range
[below −100
o
C (−148
o
F): add
0.04 % of reading for J-, K-, E-, and
N-type; and 0.08 % of reading for
T-type]
Electromagnetic
Compatibility
Susceptibility: ±2
o
C (±3.6
o
F) for 80
MHz to 200 MHz in 1.5 V/m field, for
200 MHz to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1
Maximum
Differential
Common Mode
Voltage
1 V (Maximum voltage difference
between T1 and T2)
Temperature
Scale
ITS-90
Applicable
Standards
NIST-175
Accuracy is specified for ambient temperatures
between 18
o
C (64
o
F) and 28
o
C (82
o
F) for a period of
1 year. The above specifications do not include
thermocouple error.
Replacement Parts and Accessories
Accessory Part Number
Holster and Flex Stand Assembly 1272438
AA NEDA 15A IEC LR6 batteries 376756
80PK-1 K-Type Bead Thermocouple 773135
CD-ROM 1276106
Service Manual 1276123