Everpure XC Water System User Manual


 
CARTRIDGE CHANGE PROCEDURE
E-Series Prefilter Cartridge Change
1. Shutoffpowertoequipment.Shut
off inlet water ahead of system.
2. Pushredreliefvalve.Unscrew
bowl.Removepreltercartridge.
3. Cleanbowlwithlteredwateronly;
NOSPECIALCLEANERS.
4. Feelheado-ring.Ifdry,lubricate
withhighqualitysiliconelubricant.
5. Installnewpreltercartridgein
bowl.Handtightenbowlontohead.
6. Turnonwater.Turnonpowerto
equipment.
Everpure Carbon Filter Cartridge
Change
1. Shutoffpowertoequipment.Shut
off inlet water ahead of system.
2. Openushingvalvetorelieve
pressure.
3. Hold head firmly and push upward.
Turncartridgetoleftuntilitstops.
4. Pullcartridgedownwardandoutof
head.
5. Holdheadrmly.Aligncartridge
lugwithlabel.Insertnewcartridge
firmly into head.
6. Turncartridgeright90˚until
rotation stops.
7. Repeatsteps3-6forremaining
cartridges.
Flushing
8. Withushingvalvefullyopen,open
inlet water at full force for five
minutes.Closeushingvalve.
9. Turnonpowertoequipment
PRESSURE GAUGES
Low Inlet
Pressure
Adequate Inlet
Pressure
Low Outlet
Pressure
Adequate Outlet
Pressure
SR-X Feeder Cartridge Change
1. Shutoffpowertoequipment.Shut
off inlet water ahead of system.
2. Pressredpressurereliefbuttonon
head.
3. Unscrewbowl.
4. Removeoldcartridge.
5. Feelbowlo-ring;ifdry,lubricate
withsiliconelubricant.
6. InstallnewSS-10cartridgeinto
head.
7. Handtightenbowlintohead.
8. Turnonwater.
9. Pressredpressurereliefbutton
until water comes out of valve port.
10.Turnonpowertoequipment.
For installations in Massachusetts, the Commonwealth of Massachusetts Plumbing Code 248 CMR shall be adhered to. Consult your licensed
plumber for installation of the system. This system and its installation must comply with state and local regulations.
CARTRIDGE CHANGE DETERMINATION
Replacethepreltercartridgewheneveritbecomesexcessivelydirty(observedvisuallyorbyowrestriction).Followthechangeinstructionsprovidedwith
thecartridge.Replacecarbonltercartridgeswhencapacityisreachedorwhenowbecomesinadequate,butatleastannually.Allcarbonltercartridges
onmultiplecartridgesystemsshouldbechangedatthesametime.ChangetheSS-10cartridgewhenproductlevelinthecartridgedropsbelow1/2"from
thebottomofthebowl.
ThesystempressuregaugeoroptionalEverguard
LowPressureAlarm(LPA)providesaquick,simplemeansofdeterminingwhentheltercartridge(s)
shouldbechecked.Installedonthesystemoutletmanifold,thegaugeorLPAcanbeusedtomonitorbothdynamic(owing)andstatic(line)pressure.The
cartridge(s)shouldbechangedwhenthepressuregaugeisintheredareawhileequipmentisinoperation,andyettheneedleshowsadequatelinepressure
betweencycles.Observetheneedleduringthenextequipmentoncycle.Iftheneedleregistersadequatepressure,itcanbeassumedthatthetemporary
low-pressureconditionwascausedbyabriefpowerfailureorotherincomingpressuredisruption.Ifthelowpressureconditionisnotduetoanexternal
condition,check/changetheprelterrst.Ifthelowpressureconditionisstillnotcorrected,changethecarbonltercartridges.