Bacharach ECA 450 Carbon Monoxide Alarm User Manual


 
Headquarters: 621 Hunt Valley Circle, New Kensington, PA 15068 U.S.A.
Phone: 1-800-736-4666 Canada: 1-800-328-5217 E-Mail: help@bacharach-inc.com Website: www.bacharach-inc.com
TECHNICAL DATA
Measurements & Ranges
Oxygen 0.1 to 20.9%
Carbon Monoxide 0 to 4000 ppm
(hydrogen compensated) (hydrogen compensated)
Carbon Monoxide (high)* 4001 to 80000 ppm
Nitric Oxide* 0 to 3500 ppm
Nitrogen Dioxide* 0 to 500 ppm
Sulfur Dioxide* 0 to 4000 ppm
Combustibles* 0 to 5.00% (application dependent)
Stack Temperature -4 to 2400 degrees F (-20 to 1315˚C)
Primary/Ambient Temperature -4 to 999 degrees F (-20 to 999˚C)
Pressure/Draft -27.7 to 27.7 inches of H2O
Calculations & Ranges
Combustion Efficiency 0.1 to 100.0%
Excess Air 1.0 to 250%
Carbon Dioxide (dry basis) 0 to fuel dependent maximum
NO
X
(NO
X
=NO+NO
2
) 0 to 4000 ppm
NO
X
referenced to % O
2
0 to 17000 ppm
CO referenced to % O
2
0 to 99999 ppm
NO referenced to % O
2
0 to 14900 ppm
NO
2
referenced to % O
2
0 to 2100 ppm
SO
X
referenced to % O
2
0 to 17000 ppm
Accuracy
Oxygen ± 0.3% O2 on practical concentrations of flue gas
Stack or Flue Gas Temp. ± 4°F between 32 and 255°F
(± 2°C between 0 and 124°C)
± 6°F between 257 and 480°F
(± 3°C between 125 and 249°C)
± 8°F between 482 and 752°F
(± 4°C between 250 and 400°C)
Primary-air/ambient Temp. ± 2°F between 32 and 212°F
(± 1°C between 0 and 100°C)
Pressure Draft ± 2% of reading or ± .02 in wc whichever is greater
CO ± 5% of reading or ± 10 ppm whichever is greater between
0-2000 ppm CO
± 10% of reading between 2001 to 40000 ppm CO
NO ± 5% of reading or ± 5 ppm whichever is greater between
0-2000 ppm NO
NO
2
± 5% of reading or ± 5 ppm whichever is greater between
0-500 ppm NO
2
SO
2
± 5% of reading or ± 10 ppm whichever is greater between
0-2000 ppm SO
2
HC ± 5% of full scale
Selectable Fuels Natural Gas, Oil #2, Oil #4, Oil #5, Oil #6,
Propane, Coal, Wood, Kerosene, Bagasse
Power Universal AC adapter and an internal battery pack. Adapter will
accept input voltages from 100 to 240V. Fully charged battery
pack provides a minimum of 8 hrs of operation
Pumps & Probe Two pumps are included. The first pump supplies gas sample
to the sensors. The second pump supplies fresh air to purge
the low range CO sensor when CO levels exceed 4000 ppm.
Probe includes a standard probe and hose assembly equipped
with a water trap, particulate filter, probe stop, 15 feet of hose,
and 12-inch probe tube.
Size 13.5” H x 18.5” W x 9” D
Weight 25 lbs. (11.34 kg)
Warranty 1 year, extended warranty available
* Optional
Applications
The ECA 450 is ideal for professionals concerned about combustion
efficiency, environmental compliance, or both. It enables plant
maintenance engineers and managers, industrial boiler/furnace service
technicians, energy coordinators, compliance officers, environmental
auditors and safety managers to ensure that industrial equipment is
burning efficiently while environmental regulations are being met.
The maximum operating temperature for probes included in 24-7221, 24-7224, 24-7223,
24-3024 and 24-3025 is 1472°F (800°C). The maximum operating temperature for high
temperature probes 24-3035, 24-3036 and 24-3037 is 2000°F (1093°C).
Bacharach is a registered trademark of Bacharach, Inc.
© 2003, Bacharach, Inc. All rights reserved. All information herein is subject to verification.
Product Bulletin-4138 08/03 10M DXC HPC
Printed in U.S.A.
ORDERING INFORMATION
PART NO. DESCRIPTION
24-7221 Base Unit (O
2
, CO, Ts, Ta, DP)
24-7289 Base Unit (O
2
, CO, Ts, Ta, DP) w/ wireless hand-held remote
24-7223 Heavy duty sample conditioning system (EPA/ETV Test Verified
performance). Includes Peltier chiller unit withintegral condensate pump,
15’ heated hose assembly with probe and built-in temperature controller
24-8400 NO
X
Kit with NO / NO
2
sensors and compact sample conditioner
24-8401 NO
X
and SO
X
Kit with NO, NO
X
and SO
2
sensors and compact
sample conditioner
PART NO. REPLACEMENT SENSORS
24-0788 O
2
Sensor
24-0789 CO (low) Sensor
24-0997 CO (high) Sensor
24-0881 NO Sensor
24-1055 Combustible Sensor
24-1027 NO
2
Sensor
24-0998 SO
2
Sensor (requires NO
2
option)
PART NO. MODULAR SENSORS (For Factory Installation in Base Unit)
24-3206 CO (high) Sensor
24-3207 NO Sensor
24-3209 NO
2
Sensor
24-3030 SO
2
Sensor
24-3028 Combustible Sensor
PART NO. ACCESSORIES
07-1644 Filters (pkg. of 3)
24-7059 Calibration Kit (does not include gas)
104-4027 Serial Cable DB.9M to DB.9F
24-0980 Printer Paper (5 rolls)
Distributed By: