Bacharach 24-9415 Carbon Monoxide Alarm User Manual


 
Instruction 24-9415 Bacharach, Inc. 5
Flue Gas Sample Conditioning System
Maintenance
Probe
The probe’s sponge-type lter (Figure 3, Item 9) needs to be inspected at
regular intervals for dirt contamination. Replace lter as necessary by
rst pulling out the Peltier cooler from the probe and then exchanging
the lter with a new one.
Peltier Cooler
The cooler’s wadding lter (Figure 3, Item 12) needs to be inspected at
regular intervals for dirt contamination. Replace lter as necessary by
rst pulling off the lter chamber’s top cap and then exchanging the
lter with a new one.
Condensate Pump
WARNING! The condensed water will be warm and slightly
acidic. Take all necessary precautions to prevent the water
from splashing into your eyes or contacting your skin.
Empty the pump’s reservoir (Figure 3, Item 15) by rst holding the
pump with its reservoir facing downwards. Then twist the reservoir 90
degrees (either clockwise or counterclockwise) to free it from the pump
base. Empty the reservoir; then reattach it to the pump. Before reat-
taching the reservoir, however, make sure that the reservoir’s o-ring
(Figure 3, Item 20) is properly positioned inside its grove in the pump
base.
The pump’s wadding lter (Figure 3, Item 17) needs to be inspected at
regular intervals for dirt contamination. Replace lter as necessary by
rst unscrewing the lter chamber cap and then exchanging the lter
with a new one.