3M 36 Stove User Manual


 
Operational Check
Operating sequence
17
Operating sequence
1. Turn Flow On in the setup menu of the QUESTemp° 36.
2. Plug the Air Probe into the port labeled Flow on the side of the QUESTemp°
36.
3. Turn on the Air Probe ON and make sure the green lamp is lit.
In the View or Run modes, air flow is displayed on the fifth screen.
Data Logging Air Flow
To data log air flow in the QUESTemp° 36, the following two conditions must be
met.
1. First, turn Flow On in the setup menu.
2. Second, make sure that a temperature sensor bar is connected to the Sensor
1 location (top) of the QUESTemp° 36. Flow prints out with the Sensor 1 data
therefore air flow data will only be reported if a sensor bar is plugged in. Air
flow is recorded during the run mode at the interval the QUESTemp° 36 is
setup for.
Batteries for Air Probe
The Air Probe uses a single NiMH Black&Decker VersaPak Gold battery.
Typical operating time of the battery is between 6 and 8 hours.
To change the battery, push in and twist, counterclockwise, the cap on the bottom
of the Air Probe. Pull out the battery. Insert a fully charged battery and replace the
cap.
To recharge the batteries, set the battery into the VersaPack charger. The
supplied charger accepts one or two batteries. A full charge takes 9 hours. An
indicator light shows that the battery is properly charging and it will remain on as
long as the battery is in the charger. Continuous charging is not a safety concern.
Operational Check
A verification module, 3M model 053-923, may be used to check the operation of the
QUESTemp. Remove the top sensor bar and plug the verification module into the top
of the unit. With the QUESTemp set to read in degrees Celsius, verify that the
displayed readings match those printed on the module within +/-0.5°C.
If the readings are not within the +/-0.5°C tolerance, then have the unit serviced and
calibrated.
Operational
check