Invacare Poseidon Humidifier User Manual


 
POSEIDON HUMIDIFIER
Assembly, Installation and Operating Instructions
1
OVERVIEW OF THE INVACARE
POSEIDON HUMIDIFIER
The Poseidon Humidifier helps alleviate symptoms associated
with the use of Continuous Positive Airway Pressure (CPAP)
treatment. These symptoms include a dry nose and throat. It is
intended for use with Invacare nasal CPAP systems as well as
many other manufacturers CPAP systems.
The Poseidon Humidifier is a container that is filled with water
and placed beneath the flow generator. Air from the flow gen-
erator passes over the water in the humidifier picking up mois-
ture before delivering it through your nasal cannula/mask. This
process helps to relieve irritation of the nose and throat from
use of the CPAP system.
The Poseidon Humidifier Kit Contains:
DESCRIPTION QTY.
Humidifier (blue-tinted plastic) 1
Flexible air tubing (short length) 1
Outlet Port
Inlet
Port
FILLING THE HUMIDIFIER (FIGURE 1)
1. Rinse each item of the humidifier kit in clean water and
drain completely.
WARNING
DO NOT overfill the humidifier - otherwise, in-
jury or damage may occur.
2. Holding the humidifier upright, fill the humidifier with dis-
tilled or demineralized water through either port to the level
indicated by the fill line.
FIGURE 1 - FILLING THE HUMIDIFIER
SAVE THESE INSTRUCTIONS
NOTE: Check all parts for shipping damage. If damage is found, DO
NOT use. Contact Carrier/Invacare for further instructions.
SAFETY SUMMARY
In order to ensure the safe installation, assembly and
operation of the POSEIDON HUMIDIFIER these instruc-
tions MUST be followed:
CAUTION
Federal law restricts this device to sale by or on
order of a physician, or any other practitioner
licensed by the law of the State in which he
practices to use or order the use of this device.
The Poseidon humidifier was designed to be used
with various CPAP units. ALWAYS check with your
physician/clinician to determine if it can be used
with your sleep unit/flow
generator.
GENERAL WARNINGS
DO NOT install this equipment without first read-
ing and understanding this instruction sheet AND
your CPAP owner’s manual. If you are unable to
understand these instructions, contact Invacare
technical support at the number listed on the
back page before attempting to install this equip-
ment - otherwise, injury or damage may occur.
The humidifier is for SINGLE PERSON USE ONLY.
This device is not to be used with CPAP flow
generators that have automatic pressure titra-
tion capabilities.
This device will not alter compliance parameters
when operated with a Bi-level or CPAP flow
generators.
DO NOT allow water from the humidifier to spill
through the air tubing into the flow generator as
this could result in electrical shock hazard and
could result in personal injury or equipment dam-
age. To prevent this:
DO NOT overfill the humidifier;
Disconnect the tubing at the flow generator
when not in use;
DO NOT fill the humidifier while it is
attached to the flow generator;
If water does enter the flow generator, switch
OFF the power and disconnect the flow gen-
erator. Return it to your equipment supplier
for evaluation.
Ensure that all items are properly and
securely installed and the humidifier is posi-
tioned under the flow generator.